Veterinary Monitor

Features:
1) Multi-language Operating interface selection: English, Spanish, Portuguese,
France, Arabic, and Chinese
2) Rechargeable high-energy built-in battery
3) Accurately measure the SpO2 in low perfusion situation
4) Low-power design with standby mode and no-fan type
5) Against electro surgical-interference and electric-knife; defibrillated
6) Three-application mode: monitoring, diagnosis, operating
7) Networking function with central monitoring system and software upgrade possible
8) History data storage up to 360 hours
9) Audio and visual comprehensive alarm
10) Applications in hospital like emergency pre-and post-operative care,
theater/operating room, intermediate care/step down units, labor and delivery,
hospital-based special procedure areas and outdoor ambulance surgery
11) Standard configuration: ECG, HR, PR, SpO2, Oxy CRG diagram, ST analysis, RESP,
NIBP (with venous punch), TEMP, drug dose calculation
12) Optional configuration: wireless connection, built-in printer and 2-IBP

Technical parameters:
1) ECG
2) Input: 5 wires ECG cable
3) Lead section: I,II, III/ aVR, aVL, aVF/ V, MCL1
4) Gain(mm/mV): 1/4, 1/2, 1, 2, 4
5) Frequency response: Diagnosis 0.05 - 100Hz
i. Monitor 0.5 - 40 Hz
ii. Surgery 1.0 - 25 Hz
6) Sweep speed(mm/sec): 12.5, 25, 50
7) Heart rate range: 30-254 BPM
8) Heart rate accuracy: ± 1 %
9) ST segment deviation analysis: -0.8-+0.8mV
10) ST accuracy: ±0.02mv

11) Pulse oximeter
12) SPO2 measurement range: 0 - 100 %
13) Resolution: 1 %
14) pulse measurement range: 30- 254BPM
15) Accuracy: ± 2 %
16) Resolution: 1 BPM

17) Non-invasive blood pressure
18) Measurement type : adult, pediatric, neonatal
19) Measurement range: Systolic 4.0 - 37.0 kPa
i. Diastolic 1.3 - 33.0 kPa
ii. Mean 2.6 - 35.0 kPa
20) Accuracy : ± 0.4 kPa or 5 %
21) Resolution: 0.1 kPa
22) Protection: over pressure

23) Invasive blood pressure (with venous punch)
24) Measurement range: 0~40kPa (0~300mmHg)
25) Channel: 2 channel
26) Transducer sensitivity: 5MV/V/mmHg
27) Unit display: kPa or mmHg selectable
28) Temperature (surface and rectal)
29) Measurement range: 25 - 45centigrade
30) Accuracy: ± 0.1 centigrade
31) Resolution: 0.1 centigrade
32) response time: 3 min or less than 3 min

33) Respiration Rate(RA-LL impedance and nasal cavity)
34) Measurement range: 0 - 120 BPM(adult)
35) Accuracy: ± 1 BPM or 5 %
36) Resolution: 1 BPM

37) ETCO2 (side stream type)
38) Measurement range: 0 % - 10.0 % (ins CO2 and et CO2)
39) Accuracy: ± 2 mm Hg (< 5 % Measurement value)
i. 10 % (> 5 % Measurement value)
40) Response time: 180 Ms

41) Power requirements
42) Input: 98~260 V AC, 50/60Hz
43) Consumption: 80VA or less than 80VA

44) Environment
i. Operation Storage
45) Temperature 0.5 - 40Centigrade -20 - 50 Centigrade
46) Relative humidity 80% or less than 80% 90% or less than 90%

47) Safety standard: IEC 60601-1
48) Quality System: ISO13485: 2003

49) Dimensions: 58cm x 36cm x 30cm
50) Gross weight: 8kg