Features:
1) Multi-language operating interface selection: English, Spanish, Portuguese,
France, Arabic, and Chinese
2) Rechargeable high-energy built-in battery
3) Accurately measure the SpO2 in low perfusion situation
4) Low-power design with standby mode and no-fan type
5) Against electro surgical-interference and electric-knife; defibrillated
6) Three-application mode: monitoring, diagnosis, operating
7) Networking function with central monitoring system and software upgrade possible
8) History data storage up to 360 hours
9) Audio and visual comprehensive alarm
10) applications in hospital like emergency pre-and post-operative care,
theater/operating room, intermediate care/step down units, labor and delivery,
hospital-based special procedure areas and outdoor ambulance surgery
11) Suitable for adult, pediatrics, neonates
Standard configuration: ECG, HR, PR, SpO2, Oxy CRG diagram, ST analysis, RESP*, NIBP
(with venous punch), TEMP, drug dose calculation
Optional configuration: wireless connection, built-in printer, ETCO2 and 2-IBP
Technical parameters:
1) Temperature (surface and rectal)
a) Measurement range: 25 - 45 centigrade
b) Accuracy: -1% ~1% centigrade
c) Resolution: 0.1 centigrade
d) Response time: less than 3 min.
2) ECG
Input: 5 wires ECG cable
Lead section: I, II, III/ aVR, aVL, aVF/ V, MCL1
Gain (mm/mV): 1/4, 1/2, 1, 2, 4
Frequency response:
a) Diagnosis 0.05 - 100Hz
b) Monitor: 0.5 - 40Hz
c) Surgery: 1.0 - 25Hz
Sweep speed (mm/sec): 12.5, 25, 50
Heart rate range: 30-254 BPM
Heart rate accuracy: ±1 %
ST segment deviation analysis: -0.8~0.8mV
ST accuracy: ±0.02mV
3) Respiration rate (RA-LL impedance and nasal cavity)
a) Measurement range: 0 - 120 BPM(adult)
b) Accuracy: -1~1 BPM or 5 %
c) Resolution: 1 BPM
4) Pulse oximeter
SPO2 measurement range: 0 - 100 %; resolution: 1 %
Pulse measurement range: 30- 254BPM;
accuracy: -2%~2%;
resolution: 1 BPM
5) Non-invasive blood pressure:
Measurement type: adult, pediatric, neonatal
Measurement range:
a) Systolic: 4.0 - 37.0kPa
b) Diastolic: 1.3 - 33.0kPa
c) Mean: 2.6 - 35.0kPa
Accuracy: ±0.4kPa or 5 %; resolution: 0.1kPa
Protection: over pressure
6) ETCO2 (side stream type)
Measurement range: 0 % - 10.0 % (ins CO2 and et CO2)
Accuracy: -2~2 mmHg (less than 5% measurement value);
10% (>5% measurement value)
Response time: 180ms